Collection Of theses on high power laser and plasma physics, 2016, 14 (1): 0904002, Published Online: May. 26, 2017  

Comparison Among Different Interferometers for Measuring Power Spectral Density of Optical Elements

Author Affiliations
1 中国科学院上海光学精密机械研究所高功率激光物理国家实验室, 上海 201800
2 中国科学院大学, 北京 100049
Abstract
In order to investigate the response law of different interferometers to the mid-spatial frequency errors and obtain the relatively true power spectral density (PSD) distribution, the 4D AccuFiz, ZYGO DynaFiz and ZYGO GPI interferometers are respectively used to measure the periodic ripples and scratch samples, and the comparative analysis of the measured one-dimension PSD curves is made. The results show that the responsivities to the mid- and high-spatial frequency information are different for different interferometers, and the higher the interferometer resolution is, the higher the responsivity is. The instrument transfer function (ITF) curve is used to correct the measured PSD, and the truer PSD distribution is obtained. However, in the range of some frequency bands, there still exist some difference between the ITF curve and the practical response.
References

[1] Wang D Y, English R E, Jr, Aikens D M. Implementation of ISO 10110 optics drawing standards for the National Ignition Facility[C]. SPIE, 1999, 3782: 502-508.

[2] Lawson J K, Wolfe C R, Manes K R, et al. Specification of optical components using the power spectral density function[C]. SPIE, 1995, 2536: 38-50.

[3] Spaeth M L, Manes K R, Widmayer C C, et al. National Ignition Facility wavefront requirements and optical architecture[J]. Optical Engineering, 2004, 43(12): 2854-2865.

[4] 刘红婕, 景峰, 左言磊, 等. 激光波前功率谱密度与焦斑旁瓣的关系[J]. 中国激光, 2006, 33(4): 504-508.

    Liu Hongjie, Jing Feng, Zuo Yanlei, et al. Relation between wavefront power spectral density and wings of focal spot[J]. Chinese J Lasers, 2006, 33(4): 504-508.

[5] 朱坪, 谢兴龙, 焦兆阳, 等. 大口径超短脉冲聚焦系统波前误差对时间信噪比的影响[J]. 光学学报, 2014, 34(10): 1032001.

    Zhu Ping, Xie Xinglong, Jiao Zhaoyang, et al. Influence of wave-front error on temporal signal-to-noise ratio in large aperture ultrashort pulse focusing system[J]. Acta Optica Sinica, 2014, 34(10): 1032001.

[6] Aikens D M, Wolfe C R, Lawson J K. The use of Power Spectral Density (PSD) functions in specifying optics for the National Ignition Facility[C]. SPIE, 1995, 2576: 281-292.

[7] 沈卫星, 徐德衍. 强激光光学元件表面功率谱密度函数估计[J]. 强激光与粒子束, 2000, 12(4): 392-396.

    Shen Weixing, Xu Deyan. Power spectral density function estimate in high power laser optical element surface[J]. High Power Laser and Particle Beams, 2000, 12(4): 392-396.

[8] 许乔, 顾元元, 柴林, 等. 大口径光学元件波前功率谱密度检测[J]. 光学学报, 2001, 21(3): 344-347.

    Xu Qiao, Gu Yuanyuan, Chai Lin, et al. Measurement of wavefront power spectral density of large optical components[J]. Acta Optica Sinica, 2001, 21(3): 344-347.

[9] de Groot P, de Lega X C. Interpreting interferometric height measurements using the instrument transfer function[M]//Osten W. Fringe 2005. Heidelberg: Springer Berlin Heidelberg, 2005: 30-37.

[10] Elson J M, Bennett J M. Calculation of the power spectral density from surface profile data[J]. Applied Optics, 1995, 34(1): 201-208.

[11] Youngworth R N, Gallagher B B, Stamper B L. An overview of power spectral density (PSD) calculations[C]. SPIE, 2005, 5869: 58690U.

[12] Williams W H. NIF large optics metrology software: description and algorithms[R]. United States: Lawrence Livermore National Laboratory (LLNL), 2002.

[13] Galigekere R R, Holdsworth D W, Swamy M N S, et al. Moment patterns in the radon space[J]. Optical Engineering, 2000, 39(4): 1088-1097.

[14] 程晓锋, 郑万国, 蒋晓东, 等. 用功率谱密度坍陷评价光学元件波前中频误差特性[J]. 强激光与粒子束, 2005, 17(10): 1465-1468.

    Cheng Xiaofeng, Zheng Wanguo, Jiang Xiaodong, et al. Evaluating intermediate frequency error property of wavefront of optical components with PSD collapse[J]. High Power Laser and Particle Beams, 2005, 17(10): 1465-1468.

[15] Takacs P Z, Barber S, Church E L, et al. 2D spatial frequency considerations in comparing 1D power spectral density measurements[R]. United States: Brookhaven National Laboratory (BNL), 2010: OWE5.

[16] Sykora D M, de Groot P. Instantaneous measurement Fizeau interferometer with high spatial resolution[C]. SPIE, 2011, 8126: 812610.

[17] Church E L, Vorburger T V, Wyant J C. Direct comparison of mechanical and optical measurements of the finish of precision machined optical surfaces[J]. Optical Engineering, 1985, 24(3): 243388.

[18] Takacs P Z, Li M X, Furenlid K, et al. A step-height standard for surface-profiler calibration[C]. SPIE, 1993, 1995: 65-74.

[19] Wolfe C R, Downie J D, Lawson J K. Measuring the spatial frequency transfer function of phase-measuring interferometers for laser optics[C]. SPIE, 1996, 2870: 553-557.

[20] 张蓉竹, 许乔, 顾元元, 等. 大口径光学元件检测中的主要误差及其影响[J]. 强激光与粒子束, 2001, 13(2): 133-136.

    Zhang Rongzhu, Xu Qiao, Gu Yuanyuan, et al. Testing errors and its influence of the large aperture optical elements[J]. High Power Laser and Particle Beams, 2001, 13(2): 133-136.

[21] Novak E, Ai C, Wyant J C. Transfer function characterization of laser Fizeau interferometer for high-spatial-frequency phase measurements[C]. SPIE, 1997, 3134: 114-121.

Yang Xianghui, Shen Weixing, Zhang Xuejie, Ma Xiaojun. Comparison Among Different Interferometers for Measuring Power Spectral Density of Optical Elements[J]. Collection Of theses on high power laser and plasma physics, 2016, 14(1): 0904002.