激光与光电子学进展, 2020, 57 (14): 140001, 网络出版: 2020-07-23   

纳米计算机断层扫描成像技术进展综述 下载: 2666次封面文章

Review on Development of Nano-Computed Tomography Imaging Technology
作者单位
1 天津大学精密仪器与光电子工程学院精密测试技术及仪器国家重点实验室, 天津 300072
2 天津大学南昌微技术研究院, 天津 300072
摘要
纳米计算机断层扫描(nano-CT)对于推动科技发展和生产技术进步具有极其重要的意义,在生物医学前沿研究及新材料特性研究中的作用尤为突出。概述了纳米CT成像的基本原理、分类,以及纳米CT系统的实现方式和优缺点。在具体技术细节上,重点分析实现纳米CT成像的光学元件的原理、性能、加工制造的瓶颈和应用现状。在此基础上,提出纳米CT成像系统和应用方面存在的问题,并对未来纳米CT成像的发展进行了展望。
Abstract
Nano-computed tomography (nano-CT) is of great significance for the development of science and technology and the progress of production technology, especially in frontier researches of biomedicine and new material properties. The principle and classification of nano-CT imaging, as well as the implementation, advantages, and disadvantages of the nano-CT system, are reviewed. As for the specified technology detail, the principles, properties, manufacturing bottleneck, and application status of optical components in nano-CT are described. On this basis, the existing problems in nano-CT imaging system and its applications are proposed, and the future development of nano-CT imaging is summarized.

1 引言

计算机断层扫描(CT)成像技术以其无损检测的特性在生物医学[1-3]、新材料、微系统加工与制造、地质勘探、新能源、航空航天以及考古等诸多领域有非常广泛的应用,自问世以来受到了众多关注并获得了迅速发展。

1973年Hounsfield[4]搭建了第一台真正意义上的医学CT系统。此后较长一段时间,CT系统凭借其独特的无损检测特性获得了快速发展。从第一代的单射线源单探测器的平行束CT到现在的多探测器甚至多射线源的螺旋CT[5-6],扫描效率和扫描精度都得到了很大提高。受限于时间分辨率以及低剂量的要求,目前常规的医用CT最优分辨率仅有几百微米,然而这远不能满足高新材料、生物前沿以及微系统加工与制造领域对高分辨率的需求。20世纪80年代,类似临床CT成像结构的显微CT的空间分辨率达到了147 μm×147 μm×1000 μm。随着基于FDK算法[7]的重建技术的引入,显微CT的纵向分辨率得到了极大的提高。由于X射线具有高穿透性,难以进行调制,因此CT成像系统的分辨率的提高主要依赖于投影放大,这就要求射线源具备较小的焦斑尺寸,探测器有较小的像素尺寸。受加工制造技术的限制,X射线高分辨率成像的潜力得不到有效的开发,X射线CT的系统分辨率一直处于几十微米的水平。随着微米级焦斑尺寸的射线源和高分辨率的X射线探测器(如透镜耦合X射线探测器[8])相继研制成功,基于这两者构建的显微CT的最佳分辨率一般在百纳米量级,视场为毫米量级。为了缩小焦斑尺寸,澳大利亚CRSIO公司研制了基于扫描电子显微镜(SEM)的X射线源,实现了准纳米级焦斑尺寸。利用这种准纳米级焦斑尺寸的光源结合高倍率透镜耦合探测器搭建的Nano-CT成像系统可以实现优于50 nm的分辨率[9]。然而这种基于SEM的X射线源光通量极小,清晰成像需要的曝光时间很长,成像效率极低,并不适合长时间的CT扫描。目前市面上常见的商用纳米CT大都依赖微焦斑X射线源和高分辨率成像探测器,最优的分辨率能够达到200 nm。商用纳米CT通常采用具有微纳焦斑尺寸的实验室源作为光源,这种X射线源本身亮度低,导致高分辨率的CT扫描效率低,在一定程度上制约了X射线调制元件的研制与开发。

同步辐射装置的蓬勃发展极大地促进了X射线元件的研制与发展,使得X射线的光学成像特性得以有效开发。同步辐射X射线源具有高准直、宽波段、高亮度的特点,便于利用X射线元件进行调制。随着大型同步辐射装置的进步及各式各样的X射线聚焦元件如波带片和毛细管等的开发,对X射线的聚焦与放大得以实现,使得X射线成像分辨率提高到了纳米级别,从而实现了真正意义上的纳米CT成像技术[10-13],成像分辨率在10 nm量级,其成像视野(FOV)一般只有几十微米。目前世界上多数的同步辐射研究中心都配备了高端纳米CT成像装置,包括美国先进光源(ALS)、欧洲同步辐射中心(ESRF)和日本大型同步辐射光源Spring-8,国内的中国科学技术大学国家同步辐射实验室(NSRF)、中国科学院高能物理研究所(BSRF)、中国科学院上海同步辐射光源(SSRF)。纳米CT成像系统、技术研究和应用研发逐渐走向成熟。

2 纳米CT成像原理及分类

X射线与物质之间的相互作用是十分复杂的过程,在宏观上大致可以认为X射线在物质中发生折射,则有

表 1. 纳米CT成像系统分类

Table 1. Classification of nano-CT imaging systems

PrincipleType
Imaging principleAbsorbing contrast based nano-CT
Phase contrast based nano-CT
Property of X-rayHard X-ray nano-CT
Soft X-ray nano-CT
AmplificationmechanismProjection magnifying nano-CT
X-ray microscopy nano-CT

查看所有表

n=1-δ+iβ,(1)

式中:n为X射线对样品的折射率。吸收衬度与折射率的虚部β成正比,相位衬度与折射率的实部δ成正比。当X射线穿过样品后,不仅强度发生了变化,相位也同时发生变化,在像面产生吸收衬度和相位衬度,利用这两种不同的成像机制分别形成了基于吸收衬度的CT成像技术和基于相位衬度的CT成像技术。X射线波长极短,在0.001~10 nm之间,穿透力极强,因此透射型X射线吸收衬度显微镜发展快速,种类繁多,应用更加普遍。基于吸收衬度的CT成像系统按照放大方式的不同可以分为:基于X射线投影放大型的CT成像系统和基于X射线调制元件的CT成像系统,也就是X射线显微镜。

I0为X射线的入射强度,I为X射线的出射强度。理想条件下,垂直入射的单能X射线穿过某均匀样品时满足

I=I0e-μ·Δx,(2)

式中:Δx为X射线穿过样品的厚度;μ为衰减系数。 X射线穿透样品时与其内部的物质发生一系列相互作用的结果,经验上可以表示为μ(E,Z,ρ),其中E是X射线能量,Z是样品的等效原子序数,ρ是样品密度。由于样品材质和厚度不同,出射X射线的衰减量也不同,从而可以在探测器上获得包含样品内部结构的投影图像[14]

CT成像系统的主要部件包括:X射线源、X射线探测器、调制元件、样品台及运动控制系统。扫描成像的过程是:通过X射线调制元件对射线进行聚焦后作为照明光源, X射线透过X射线调制元件进行投影放大,样品旋转360°,通过探测器采集到包含足够信息量的吸收投影图像或相衬投影图像,通过三维重建获得样品内部的三维信息。三维重建的过程可以简单理解为求取样品内部各离散点的衰减系数μ(E,Z,ρ)的过程。利用重建方法得到的某一个断层的二维截面,之后可以将所有的二维断层截面拼合起来,得到样品的三维图像。重建算法可以分为解析类重建算法和迭代类重建算法。解析类重建算法的重建速度较快,对于完整的数据能够得到高质量的重建图像,缺点在于要求扫描角度多、数据统计涨落小。迭代类重建算法适用于各种扫描模式,对扫描角度小、统计涨落大的数据重建得到的图像优于解析类算法,缺点是重建速度慢。典型的解析类重建算法有傅里叶变换重建算法、滤波反投影算法和反投影滤波重建算法;典型的迭代类重建算法有ART重建算法、Richardson重建算法等[14]

纳米CT成像系统能够实现优于100 nm的分辨率,其成像系统结构精密复杂,有多种分类方式。按照成像机制的不同,纳米CT成像系统可分为基于吸收衬度的纳米CT成像系统和基于相位衬度的纳米CT成像系统;按照X射线性质的不同,纳米CT成像系统可以分为硬X射线CT成像系统和软X射线CT成像系统。依据系统结构的不同,纳米CT成像系统又可以分为基于投影放大型的CT成像系统和X射线显微镜型的纳米CT 成像系统。表1对X射线CT成像系统的分类进行了汇总。基于投影放大型的CT成像系统结构原理简单,是最经典的结构类型,通常这种类型只能采用吸收衬度成像的方式。X射线显微镜型的纳米CT 成像系统根据成像机制的不同可以分为基于吸收衬度的纳米CT成像系统和基于相位衬度的纳米CT成像系统。通常情况下包含相移元件的系统为基于相位衬度成像机制的纳米CT成像系统。

基于吸收衬度的纳米CT成像系统在结构及重建算法方面与传统CT成像系统相似,成像装置结构简单,重建方法成熟,鲁棒性好,适用于对金属类、骨骼牙齿、芯片等原子序数较大的物质成像。然而,原子序数较小的生物软组织样本,对绝大多数波段范围的X射线的吸收能力相差不大,利用这种成像技术难以区分各种不同的轻质物质。

针对生物软组织如细胞等,基于吸收成像的水窗软X射线纳米CT成像有独特的优势。水窗软X射线是指能量位于284~543 eV的射线,在这个波段范围内,含碳的有机物对X射线的吸收是水对X射线的吸收量的10倍以上,对于大多数含水的有机物如细胞等,只需利用吸收成像便能达到很高的对比度。当然软X射线波长较大,景深小,成像范围受到严重限制,同时长时间的辐照对生物活体样品会造成损伤。

基于相衬效应的纳米CT成像技术是吸收型纳米CT成像技术的延伸和发展,对轻量元素组成的样品如生物软组织进行成像有独特的优势。这类样品所含元素的原子序数较小,在硬X射线波段内相移截面比吸收截面大2~3个数量级,因此利用相衬成像机制进行成像能获得更高的图像衬度以及更高的探测灵敏度。同时,硬X射线的相衬成像景深较大,约为几十微米,这样的景深可以完整成像大多数真核细胞[15-21],因而针对轻质材料,基于相衬成像机制的CT成像技术更适合。国家同步辐射实验室利用硬X射线显微术和纳米断层扫描在细胞成像方面的独特优势,开展了相关方面的研究,弥补了光学显微镜和电子显微镜在细胞成像方面的不足,这在国际上也是领先的探索,为将纳米CT成像技术发展成为一种有力的细胞成像工具做出了出色的工作[22]

3 基于投影放大的纳米CT成像技术

基于投影放大的纳米CT成像技术主要通过几何投影放大直接进行成像,通过使用具有纳米焦斑尺寸的射线源和较大的几何放大比(源像距与源物距之比)达到纳米级别的分辨能力(图1)。因而,直接投影放大的纳米CT成像技术还依赖于具有较小像素尺寸的X射线探测器。

根据CT成像系统的有效射束宽度公式[14][(3)式],成像分辨率经验上为有效射束宽度的一半,由焦斑尺寸S、探测器的像元尺寸d和系统的几何放大比M共同决定:

BW=d2+[S(M-1)]2M(3)

射线源到待测样品的距离Z1越小,待测样品到探测器的距离Z2越大,则投影放大比M=(Z2 +Z1)/ Z1越大,越能够分辨更小的细节。焦斑尺寸S具有一定的尺寸,在投影过程中不可避免地会出现弥散斑,散斑尺寸为S',仅增加几何放大比并不能有效地提高分辨率。如何通过技术手段减小焦斑尺寸和像元尺寸成为纳米CT发展的重要方向,而这两者都依赖于X射线元件的研制。一方面,通过X射线元件能够实现对X射线的聚焦,从而实现焦斑尺寸的减小;另一方面,X射线聚焦元件能够像普通光学透镜一样实现光学放大,从而降低了对探测器像元尺寸的要求。

图 1. 投影放大成像原理图

Fig. 1. Principle of projection magnification imaging

下载图片 查看所有图片

3.1 纳焦斑X射线源的获取

通常情况下,X射线源包括实验室X射线源和同步辐射X射线源两种。实验室X射线源亮度低,且焦斑尺寸越小,光通量越小,则成像效率越低。同时焦斑尺寸越小,X射线源的造价越昂贵,因此目前市面上鲜见具有纳米焦斑尺寸的商用X射线源。一种纳焦点射线源的实现方式是:将数十纳米的金属靶安置到改造扫描电子显微镜(SEM)源的真空腔内,电子束轰击金属靶,从而产生数十纳米焦斑尺寸的X射线源[23]。另一种是利用X射线元件聚焦实现纳焦斑,例如同步辐射X射线源配合X射线调制元件形成纳米焦点的射线源。

X射线聚焦元件按照原理可分为四大类:衍射型、反射型、折射型以及吸收型。表2列举了常见X射线聚焦元件的种类名称及能够实现的最小束斑直径。由于X射线有极强的穿透性,X射线元件的材质一般为金属,如铝或铍。同时由于硬X射线(能量大于1 keV的X射线)和软X射线对物质的穿透能力不同,所采用的元件也不同。波带片是最典型的衍射型X射线聚焦元件,能够实现较高的分辨率,穿透率较高,多用作较低能X射线聚焦元件。利用交错堆叠技术加工的波带片实现了7.8 nm的焦斑尺寸,代表了目前世界最高水平[24] 。椭球聚焦镜的传输效率高达90%以上,多作为较高能量的X射线(通常指能量大于8 keV)的聚焦元件。ESRF等大型同步辐射中心开展了诸多实验探究,利用聚焦椭球镜的纳米CT实现技术,得到数十纳米的焦斑尺寸。德国维尔茨堡大学利用X光波导实现了对具有10 nm×10 nm焦斑尺寸的X射线发散光束[25]的调制,这标志着用X射线可以实现准纳米级的探测。中国科学院上海应用物理研究所依托SSRF自主设计研发的单毛细管椭球镜能够将能量为9 keV的X射线聚焦到40 μm,并据此实现纳米成像[26]。澳大利亚CRSIO公司通过改造SEM源创造出具有数百纳米焦斑的X射线源,如图2所示,利用这种射线源搭建的纳米CT成像系统实现了35 nm的分辨率[23]

图 2. 利用改制的SEM装置实现纳米CT成像的原理图[9]。(a) X射线点光源投影放大光路图; (b)改造SEM获得纳米光源的主要构造及部件;(c)锥束CT成像系统的几何原理;(d)断层数据集的虚拟截面

Fig. 2. Schematic diagram of nano-CT imaging using modified SEM device[9]. (a) Optical path diagram of X-ray point source projection magnification; (b) main structure and components of modified SEM for obtaining nano-light source; (c) geometric principle of cone-beam CT imaging system; (d) virtual cross section of tomographic dataset

下载图片 查看所有图片

基于投影放大原理实现的纳米CT成像原理简单,但是对硬件的要求非常苛刻,极度依赖于具有纳米焦斑尺寸的X射线源。大部分X射线聚焦元件设计精密,加工困难,调制效率低,其研制与开发有赖于具有高通量的大型同步辐射装置。

3.2 高分辨率X射线探测器

高分辨率X射线探测器有光学透镜耦合型探测器和光纤型探测器两种,相较于普通的X射线平板探测器具有分辨率高的特点。光纤耦合CCD探测器结构如图3所示,X射线经过闪烁体转化为可见光之后,通过光纤耦合到CCD探测器上,光纤耦合的方法可以有效地增大CCD探测器的有效探测面积,同时提高了分辨率。但是,在成像过程中,CCD探测器在X射线的直通光路中,高能量的X射线照射会对CCD造成一定程度的辐射损伤。

图 3. 光纤耦合探测器结构示意图[27]

Fig. 3. Structural diagram of coupling of fiber and detector[27]

下载图片 查看所有图片

透镜耦合探测器是纳米CT成像系统的重要组成部分,这类探测器由闪烁体、可见光放大光路及CCD构成[28-30](图4)。根据有效射束宽度的公式[(1)式],CT成像系统的分辨率由X射线源焦斑尺寸和探测器像素尺寸决定。系统放大倍数为X射线投影子系统的放大倍数和光学子系统放大倍数的乘积。理论上,成像分辨率可以获得一个量级的提升。受光学衍射极限的限制,单独使用这种方法仅能达到几百纳米的分辨率。因此,通过X射线放大提高CT系统分辨率是必然的选择。

图 4. 透镜耦合探测器的结构示意图[29]

Fig. 4. Structural diagram of coupling of lens and detector[29]

下载图片 查看所有图片

3.3 样品台及运动控制平台

纳米CT采用的机械运动系统比显微CT的运动控制系统精度更高,但二者没有明显的界限。当前商用的线性运动位移台和旋转台的精度在数十到几百纳米不等[31],通常人们会采用各种外部方法检测运动误差并对其进行校准的方法来确保运动精度。2006年,Nicola等[32]利用电容式位移传感器在气浮转台上对转台的转动误差进行了校准,将转台误差降低到10 nm以下。Kim等[33]在2013年利用激光干涉仪作为距离传感器成功将转台的径向误差校准到40 nm以内。2014年Xu等[34]利用5个电容式位移传感器和一个标准件,通过重复实验,成功测出了转台转动时的系统误差,包括端跳、径跳、偏摆等。2017年Stankevi [31]将光纤干涉仪作为传感元件、标准圆柱作为参考元件,并结合稳健的校正算法构建了纳米定位系统,将偏摆误差从1.25 μm下降到40 nm。 国内在纳米CT方向的研究起步较晚,对于样品旋转台机械参数的研究并不系统,天津大学的Zhao等[35]利用高精度的电容传感器以及转轴误差分析仪监测扫描过程中的运动误差,提出相应的转台误差校准理想数学模型,并根据实验结果对其进行修正,将旋转误差控制在数十纳米以内[35]。Fu等[36-37]利用图像处理的方法对CT系统的投影图像机械误差进行检测并提出校正方法。该方法可用于动态几何参数的检测,对于由几何参数引起的伪影校正取得了一定的效果。机械部件的稳定性研究对于提高纳米CT成像系统这种长时间扫描成像系统的分辨率有极其重要的意义。

4 基于X射线放大的纳米CT成像技术

图5(a)~(c)分别为可见光显微镜成像原理图、软X射线显微镜成像原理图以及硬X射线显微镜成像原理图。基于X射线放大技术实现的纳米CT成像是指利用X射线元件搭建X射线显微镜,实现X射线的显微放大,从而得到纳米级分辨率的图像,如图5 (b)和图5(c)所示。X射线是光的一种,其波段范围是0.001~10 nm,按照光学的衍射极限公式, X射线光学显微镜理论上的极限分辨率至少应在0.001 nm量级,而目前纳米CT成像的分辨率仅能突破10 nm,X射线在纳米成像检测领域仍有巨大的发展潜力。实现X射线显微镜系统的关键在于X射线调制元件的研制,表2汇总了当前大多数X射线元件及其重要参数。

图 5. 光学显微镜与X射线显微镜对照图。(a)可见光显微镜成像原理图;(b)软X射线显微镜成像原理图;(c)硬X射线显微镜成像原理图

Fig. 5. Comparison of optical microscope and X-ray microscope. (a) Imaging schematic of visible light microscope; (b) imaging schematic of soft X-ray microscopy; (c) imaging schematic of hard X-ray microscope

下载图片 查看所有图片

4.1 X射线波带片

波带片被广泛用作软X射线波段的聚焦元件[24,38-45,57,73-80],利用波带片搭建的纳米CT成像系统如图6所示。成像分辨率由波带片最外环的宽度决定,调制效率由厚度决定。为了兼顾分辨率和成像效率,通常期望波带片具有更高的宽深比。目前已有诸多研究机构已经开展大宽深比波带片的研究工作,其中美国加州大学伯克利国家实验室利用交错技术成功制造了最外径宽度为15 nm的波带片,利用一个聚焦波带片和一个显微波带片,在软X射线波段(250 eV~1.8 keV)实现了15 nm的分辨率[39]

图 6. 利用波带片搭建X射线纳米CT成像系统[39]

Fig. 6. X-ray nano CT imaging system with zone plate[39]

下载图片 查看所有图片

波带片具有较高的聚焦效率,是典型的X射线调制元件,各研究小组开展了诸多工作以研究扩大其适用范围的方法。瑞士保罗谢勒研究所利用基板双面叠加技术加工制造了外径宽度为30 nm的波带片,利用能量为9 keV的硬X射线波段实现了30 nm的分辨率,效率达到了9.9%[24] 。Li等[40]结合金属辅助化学蚀刻和原子层沉积等最新技术在15 μm厚的硅晶片上制造出了厚度为8 μm、最小带宽为16 nm的铂波带片,宽高比达500∶1,对于20 keV的X射线的衍射效率为4.8%。这个数值比大多数菲涅耳波带板更高,对于如此高能量的X射线,其效率值接近用多层镀膜Kirkpatrick-Baez反射镜和多层劳厄透镜实现的效率值。美国阿贡(Argonne)国家实验室提出了基于平面镀膜制作技术的X射线多层膜劳厄透镜(multilayer Laue lens),其结构如图7所示[48-49],它首先镀制足够厚的内层薄膜,生长各薄膜层的厚度可以控制到埃量级,从而可以实现对相当于波带片厚度的各层膜厚的精确控制,两个多层膜劳厄镜的倾角决定其聚焦性能。目前诸多研究机构开展了此方面的研究[50-54, 81],对12 keV的硬X射线实现了11.2 nm的聚焦[52]。这种多层膜劳厄透镜的引入相当于增加了波带片的宽深比,实现了更高的调制效率,然而这种技术受到镀膜厚度的限制。

图 7. 波带片结构示意图。(a)经典波带片结构示意图;(b)多层膜劳厄镜结构示意图[48]

Fig. 7. Structural diagram of zone plate. (a) Structural diagram of classical zone plate; (b) structural diagram of multilayer film Laue mirror[48]

下载图片 查看所有图片

表 2. 常用X射线聚焦元件分类与重要参数汇总

Table 2. Classification and important parameters of common X-ray focusing components

CategoryComponentBeam diameter /nmEnergy /keVEfficiency /%Ref. No
DiffractingFresnel zone plates35810-70[24,38-45]
Multilayer zone plates513.810-20[13,46-47]
Multilayer Laue lenses1014.610-30[48-54]
AbsorbingPinholes/slits100011[55-56]
ReflectingCapillaries50880[12,57-59]
Waveguides1015<25[25,60]
Kirkpatrick-Baez mirrors720>50[52,61-63]
RefractingSpherical compound refractive lenses10008several 10%[11, 63-64]
Parabolic compound refractive lenses5021>60[26,65-70]
Kinoform lenses1008>50[71-72]

查看所有表

4.2 复合折射透镜

复合折射透镜是另一种常用的X射线聚焦元件,由多个特定形状的X射线凹透镜组合而成,主要分为球形复合透镜[图8(b)]、 Kinnoform透镜、抛物面复合透镜及平面折射透镜[图8(a)]等类型。其聚焦原理与可见光透镜的聚焦原理相似,作为一种能量波,X射线穿透物质时既存在折射现象又有衰减现象,其折射特性决定了透镜对X射线的聚焦性能。但是由于其折射系数非常小(典型数值在10-7~10-5之间)且衰减特性强,X射线复合折射透镜经历了漫长的发展过程。真正意义上的X射线复合折射透镜在1996年问世,Snigirev等[11]在一块铝板上加工了30个直径为600 μm的圆孔,将能量为14 keV的X射线聚焦在1.8 m的焦距位置,获得的光斑尺寸为8 μm。这种球形的复合折射透镜存在明显的球差,严重影响成像质量[82]。抛物面复合折射透镜的曲面为抛物面[70,83-84],能在二维方向上对X射线进行聚焦,能够克服像差,但是抛物面的加工难度大。平面复合折射透镜就是一个抛物面复合折射透镜的切面,能产生一维方向的聚焦[63, 85]。随着集成电路制造技术和光刻技术的发展,从某种程度上来说抛物复合折射透镜的加工难度得到降低,因此平面复合折射透镜得到了发展。Kinnoform透镜通过切掉对聚焦没有作用的部分,减小透镜对X射线的衰减作用[71],效率可以达到90%[69],但是目前并没有成熟稳定的加工途径,很多都是在研究阶段[72,86]。由于复合折射透镜通常厚度较大,多采用轻质材料,如硅单质、光刻胶、金刚石、玻璃碳及金属铍、铝、锗等,而对于高能量波段(大于150 keV)的X射线,金属镍是不错的选择[87]

图 8. 复合透镜结构示意图。(a)平面复合折射透镜;(b)球形复合折射透镜[91]

Fig. 8. Structural diagram of composite lens. (a) Planar composite refractive lens; (b) spherical composite refractive lens[91]

下载图片 查看所有图片

目前应用最成熟的是抛物面复合折射透镜,其适用能量区间广、安装简易,成为同步辐射实验站聚焦元件的首选。受复合折射透镜单位透镜表面的缺陷和堆叠精度的限制,这类X射线元件存在严重的球差,是X射线成像领域内的一大研究热点。越来越多的研究人员试图从光学的角度来理解和描述X射线,Seiboth等[88]对X射线透镜的波前误差进行了测定,并开展了校正工作,希望通过X射线的光学函数来评价和分析X射线成像过程,进一步了解X射线特性[89-90],从而能够从理论上对X射线断层成像提出进一步的发展方向。

4.3 布拉格放大器

X射线布拉格放大器是一种由两块正交放置的具有Kirkpatrick-Baez几何形状的硅晶体组成的具有X射线聚焦功能的元件,结构如图9所示。1948年,Kirkpatrick和Baez[61]将反射镜(K-B透镜)用于X射线成像系统,将两个维度的放大器垂直放置,实现图像保真放大。由于X射线穿透力强,X射线反射镜的反射率低,只有在入射角小于临界角的情况下才能得到较高的反射率。为解决这个问题,研究人员采用了在反射面镀膜的方法。按照薄膜层数的不同,X射线反射镜一般分为两类。一类是单层膜全反射镜,用来以非常小的入射角反射较宽能带的光束;另一类是多层膜反射镜,其允许以更大的入射角反射光束,但其工作的能带较窄。布拉格放大器的优势在于能在硬X射线波段实现较高效率的调制,且通过调节K-B透镜的曲面程度调节放大倍数[30,92-95] 。其局限性在于安装精度要求高,K-B透镜的面型误差难以控制。反射光学器件具有固有的消色差特性,适合在宽能带X射线实验中应用,如图10所示。对于能量大于20 keV波段的硬X射线能够实现20×20~100×100等放大比的成像,最高分辨率可达140 nm[96] 。基于原子尺度的超平滑抛光技术和自适应光学,目前K-B镜在20 keV的能量下获得的最小聚焦光斑为7 nm[62,93]

图 9. 椭圆面型布拉格放大器示意图[95]

Fig. 9. Schematic diagram of elliptical type Bragg amplifier[95]

下载图片 查看所有图片

图 10. X射线布拉格放大器的组成结构示意图[96]

Fig. 10. Diagram of structure of X-ray Bragg amplifier[96]

下载图片 查看所有图片

4.4 X射线毛细管

X射线毛细管是根据内全反射原理而制成的光学器件,基本结构如图5(c)所示。当X射线以掠入射角入射到空心玻璃毛细管中时可以达到很高的反射率[97]。特殊加工的毛细管利用这种全反射特性,通过改变 X 射线的传播方向对X射线进行调制,从而实现 X 射线的聚焦或准直。准直毛细管光学透镜可将锥束X射线调制为均匀出射的准平行X射线束,具有平行度高、射线强度强等优点。聚焦毛细管顾名思义能够实现X射线的聚焦,通常可以实现大于90%的调制效率,因此可以极大地缩短扫描时间,同时能够实现X射线“透镜”的作用,达到对X射线成像放大的效果。

根据光通道数量的不同,可以将毛细管分为单毛细管和多毛细管。单毛细管又被称为椭球镜,具有聚焦光斑小、增益高、体积小等优点,大多用作硬X射线纳米CT的调制元件。其结构原理如图11所示,X射束能够通过一个单毛细管光滑内壁发生全反射,从而实现光束聚焦,对于能量为10 keV的X射线,临界角为0.2°。目前用于全视场纳米CT成像的单毛细管被国外商业公司垄断,日本同步辐射研究所采用冲压技术刻划抛物线槽,进一步提高了调制效率[66]。如何控制面型误差是提高椭球镜性能参数的关键,上海同步辐射中心在单毛细管椭球镜的制造方面进行了理论研究和仿真分析,由于制造技术的局限性,实际的加工制造仍旧困难重重[26, 98]。目前加工水平难以避免透镜椭球面的形状误差,成为研制椭球镜元件的瓶颈。

图 11. 毛细管聚焦原理示意图

Fig. 11. Schematic diagram of capillary focusing

下载图片 查看所有图片

多毛细管是由一簇横截面呈六方形状的空心玻璃纤维管紧密排列而组成的毛细管阵列,如图12所示,它接收面大,X射线利用率高,但聚焦光斑较大。毛细管聚焦在同步辐射X射线源、实验室X射线源的聚焦和各种X射线成像中都有广泛的应用。然而,针对毛细管元件的诸多研究工作更多的是集中到X射线荧光CT成像[99-101]、X射线衍射成像[102-103]等,在X射线CT成像领域的应用极为有限[104]

图 12. 单毛细管聚焦示意图(左)和多毛细管聚焦示意图(右)

Fig. 12. Single capillary focusing diagram (left) and multi-capillary focusing diagram (right)

下载图片 查看所有图片

5 基于相位衬度的纳米CT成像技术

基于相位衬度的纳米CT成像技术是吸收型纳米CT成像技术的延伸和发展,对轻量元素组成的样品如生物软组织进行成像有独特的优势。轻量元素组成的样品所含元素的原子序数较小,在硬X射线波段内相移截面比吸收截面大2~3个数量级,因此利用相衬成像机制进行成像能获得更高的图像衬度以及更高的探测灵敏度,这类成像技术比较常见的有泽尼克相衬成像、光栅相衬成像及Ptychofraphic(简称PTY)成像技术。其中泽尼克相衬成像和PTY成像技术对X射线的空间相干性要求极高,这对其发展造成了一定的局限性。另外一个非常关键的问题是相位信息的解析重构,目前主流的做法是引入微分相位衬度成像机制,以解决部分相位衬度定量化和相位周期性问题。相衬成像技术的成像机制决定了其抗干扰能力差,对系统装配精度和环境的要求很高。

X射线泽尼克相衬成像技术是泽尼克相衬显微镜在X射线波段的延伸,泽尼克相移环置于物镜波带片后焦面,使样品直通光产生π/2或3π/2的相移,进而与样品衍射光在探测器上发生干涉,将相位信息转换为强度信息被探测器探测到。同时这种成像技术有很高的灵敏度,是一种高分辨率、高衬度的成像方法[16,18,105-106],在细胞、生物组织以及有机材料成像中有着广泛的应用,成为生命科学、材料科学、信息科学、智能芯片、能源科学和纳米科技等热门研究领域不可或缺的研究手段[21,107-117]

2003年Momose等[118]将X射线泰保-劳(Talbot-Lau)光栅干涉仪引入硬X射线CT成像系统,使X射线相位衬度成像技术的光源从同步辐射光源扩展到了常规光源。基于投影条纹的X射线光栅相衬成像方法对相干性要求低,光栅周期变大后可以降低光栅的制造难度,有望获得更大的成像视场以及更高的成像能量,在临床医学和工业扫描中难以避免受到震动和系统部件安装误差的影响,Hauke等[119]提出了一种增强重建算法来调整光栅的位置,从而补偿这些误差。

PTY成像技术在国内被翻译为层叠成像[120],成像原理是通过改变照明光束和样品的相对位置,获得系列交叠区域的衍射图样,恢复得到样品的振幅分布。X射线PTY既能充分发挥X射线的高穿透力,又能实现无透镜成像的高敏感度相衬成像[120-123]。生物组织样本、聚合物或纤维复合材料等轻质样品对X射线的吸收较弱,采用相位敏感成像的方法能够有效提高衬度,而常规的X射线吸收成像方法并不适用。层叠成像技术在大幅面成像和高分辨率成像方面具有巨大潜力。与传统的衍射成像相比,此成像系统无需透镜,不但避免了透镜的像差对成像恢复质量的影响,而且也消除了数值孔径对样品尺寸的限制。该技术不仅可以与多切片技术结合,消除样品的厚度效应,还可以与X射线荧光结合。但目前该技术依然存在着一些问题,比如样品太厚导致投影近似值的不确定性,最突出的问题在于其成像效率与质量之间的矛盾。

相比于传统的X射线吸收机制的成像技术,相衬成像技术仅经历了短短30年的发展,在成像装置和成像算法方面距离发展成熟尚有一段距离。然而,相衬成像技术对轻质材料内部成像有巨大优势,成为极有潜力的发展方向之一。

6 纳米CT成像现存问题

6.1 纳米CT向计量型的转化问题

对于具有内部特征的样品的尺寸测量,CT是少数适用的测量工具,随着CT成像系统相关标准的发展,CT已逐渐发展为计量工具。虽然目前纳米CT的最优分辨率已经突破10 nm,但是仅限于定性测量,还不能实现定量测量的功能。生物医学、高新材料、先进制造的快速发展对CT成像提出了从定性测量到定量测量的要求,成为纳米CT的主流发展方向之一[124-128]。然而,作为测量仪器,CT设备的校准规范及流程目前在国际上尚未达成共识,德国、美国、中国等国家发布了各自的标准,但并未形成ISO标准;针对CT测量的发展,德国Zeiss公司、Werth公司及日本的Nikon公司分别研制了部分计量型CT产品,但分辨率仅在数微米,纳米CT的计量功能的实现还需要大量的研究。

6.2 系统几何参数和机械定位精度

CT成像的重构过程是建立在理想几何位姿的基础上,然而实际上难以避免系统随机震动和装配误差给纳米CT的实现造成困难。对于利用同步辐射X射线源的纳米CT,X射线的获取和调制所需要的部件繁杂,进一步给调试和校正工作带来困难。采用实验室源的纳米CT系统则难以避免射线源的震动。除此以外,扫描过程中运动系统的震动会造成随机干扰。受机械加工水平的限制,旋转扫描模式还会造成扫描中心的偏摆。通常纳米CT相较于显微CT时间更长,对系统稳定性和系统部件的几何位置的精准度比显微CT有更加严苛的要求,然而目前对这方面的研究资料却不多见。

6.3 环境温度对成像的影响

纳米CT成像设备属于封闭式空间设备,在样品扫描过程中,射线源连续发射X射线,会引起内部环境温度的变化,热膨胀的存在使部件位置关系发生变化,造成图像分辨率下降。因此,在纳米CT成像设备中,需增加控温设备或对各部件位置进行实时监控。且对于采用实验室源的纳米CT,由于大多数纳米X射线源光强较小,扫描时间大大延长,普通探测器不可避免地会出现噪声增强的现象[129],对探测器暗电流的检测和校正对于提高系统的分辨率不容忽视[130]

7 结束语

纳米CT成像系统在生物医学、微系统制造、材料学等多个研究领域已经成为与扫描电子显微镜、扫描隧道显微镜等一样不可或缺的成像仪器,目前已经能够实现亚细胞层级的三维成像。是最重要的无损检测手段之一。依托于同步辐射技术的进步,纳米CT成像技术获得了快速发展。世界上绝大部分的同步辐射装置均设置有CT成像系统,国际上的大型同步辐射装置主要分布在北美(ALS、APS、SLAC、CHESS、CLS等)、欧洲(ESRF、BESSY、SLS等)、东南亚(日本Springer-8、新加坡SSLS、韩国)及澳大利亚(AS)等地区。目前国内BSRF、SSRF、国家同步辐射实验室也在纳米CT成像领域做出了突出的贡献。研究人员基于SSRF在X射线聚焦元器件的制造领域进行了深入研究。国家同步辐射实验室建有中国第一台以真空紫外和软X射线为主的专用同步辐射光源。基于软X射线的纳米CT在纳米材料、细胞科学、多孔电极材料等领域中的应用研究取得了很多进展,填补了国内相关领域的空白[131]。理论方面,朱佩平等[132]在以吸收成像为机制的传统纳米CT中,引入折射和散射光强响应成像机制,提出具有三种定量成像机制的纳米CT三维成像方法,实现对直径5 μm以上细胞的完整三维成像,分辨率达到纳米量级。

纳米CT成像技术分辨率的提升依赖于X射线元件的发展。突破制造水平对促进纳米CT的研究和发展有重要的意义。同时,这种严苛的加工制造要求带来的高昂的经济成本也是纳米CT成像技术面临的困难之一。

X射线归根结底是一种光波,基于可见光光学成像的原理也可推广到X射线成像,得益于同步辐射装置和各种X射线元件的开发,纳米CT技术获得了快速发展,克服了光学显微镜的分辨率低和电子显微镜穿透性差的局限性,然而目前数十纳米的分辨率远未达到X射线的分辨率极限,其分辨率和稳定性还有较大的提升空间。

纳米CT是一种极其实用且拥有巨大应用前景的无损检测手段,能够实现定量检测是众多应用领域的需求,也是开发仪器的科研人员必然的努力方向之一。目前纳米CT领域的研究热点在于进一步提高系统的分辨率和提高系统测量的准确度。前者依赖于X射线成像元件的开发,后者则要求将成像系统的不稳定性降至足够小的范围。提高测量准确度的手段有许多工作可以开展,例如监测射线源的不稳定性曲线、机械系统的振动和噪声曲线等,对投影数据的调整有利于提高重建图像的质量,探究X射线与样品相互作用的热效应对尺寸测量的影响,这些工作的开展均有助于获取更加准确的图像,从而进一步实现溯源与计量。

参考文献

[1] 颜功兴, 刘占芳, 冯晓伟. 基于CT扫描和激光烧结技术的上颌骨及牙列三维仿真与修复[J]. 中国激光, 2009, 36(10): 2538-2542.

    Yan G X, Liu Z F, Feng X W. Three dimensional simulation and repair of skull maxilla and dentition based on CT scanning and laser sintering technologies[J]. Chinese Journal of Lasers, 2009, 36(10): 2538-2542.

[2] 张朝霞, 陈晓冬, 单建丰, 等. 基于多层螺旋CT血管分析的感兴趣冠脉段最佳造影角度计算[J]. 中国激光, 2011, 38(11): 1104003.

    Zhang X X, Chen X D, Shan J F, et al. Calculation of optimal angiographic angle for segment of interest based on multislice computed tomography vessel analysis[J]. Chinese Journal of Lasers, 2011, 38(11): 1104003.

[3] 苗光, 李朝锋. 二维和三维卷积神经网络相结合的CT图像肺结节检测方法[J]. 激光与光电子学进展, 2018, 55(5): 051006.

    Miao G, Li C F. Detection of pulmonary nodules CT images combined with two-dimensional and three-dimensional convolution neural networks[J]. Laser & Optoelectronics Progress, 2018, 55(5): 051006.

[4] Hounsfield G N. Computerized transverse axial scanning (tomography): Part 1. Description of system[J]. The British Journal of Radiology, 1973, 46(552): 1016-1022.

[5] Taguchi K, Aradate H. Algorithm for image reconstruction in multi-slice helical CT[J]. Medical Physics, 1998, 25(4): 550-561.

[6] Kalender W A. Thin-section three-dimensional spiral CT: is isotropic imaging possible?[J]. Radiology, 1995, 197(3): 578-580.

[7] Feldkamp L A, Davis L C, Kress J W. Practical cone-beam algorithm[J]. Journal of the Optical Society of America A, 1984, 1(6): 612-619.

[8] Martin T, Koch A. Recent developments in X-ray imaging with micrometer spatial resolution[J]. Journal of Synchrotron Radiation, 2006, 13(2): 180-194.

[9] Brownlow L, Mayo S, Miller P, et al. Towards 50-nanometre resolution with an SEM-hosted X-ray microscope[J]. Microscopy and Analysis, 2006, 20(2): 13-15.

[10] Bilderback D H, Thiel D J, Pahl R, et al. X-ray applications with glass-capillary optics[J]. Journal of Synchrotron Radiation, 1994, 1(1): 37-42.

[11] Snigirev A, Kohn V, Snigireva I, et al. A compound refractive lens for focusing high-energy X-rays[J]. Nature, 1996, 384(6604): 49-51.

[12] Bilderback D, Hoffman S, Thiel D. Nanometer spatial resolution achieved in hard X-ray imaging and Laue diffraction experiments[J]. Science, 1994, 263(5144): 201-203.

[13] Kamijo N, Tamura S, Suzuki Y, et al. Fabrication and testing of hard X-ray sputtered-sliced zone plate[J]. Review of Scientific Instruments, 1995, 66(2): 2132-2134.

[14] 张朝宗, 郭志平, 张朋. 工业CT技术和原理[M]. 北京: 科学出版社, 2009.

    Zhang CZ, Guo ZP, ZhangP. Technology and principles of industrial CT[M]. Beijing: China Science Publishing & Media, Ltd., 2009.

[15] Kosior E, Bohic S, Suhonen H, et al. Absolute zinc quantification at the sub-cellular level by combined use of hard X-ray fluorescence and phase contrast imaging techniques[J]. Journal of Physics: Conference Series, 2013, 463: 012021.

[16] Lim J, Kim H, Park S Y. Hard X-ray nanotomography beamline 7C XNI at PLS-II[J]. Journal of Synchrotron Radiation, 2014, 21(4): 827-831.

[17] Takeuchi A, Suzuki Y, Uesugi K. Development of scanning-imaging X-ray microscope for quantitative three-dimensional phase contrast microimaging[J]. Journal of Physics: Conference Series, 2013, 463: 012034.

[18] Takeuchi A, Uesugi K, Suzuki Y. Zernike phase-contrast X-ray microscope with pseudo-Kohler illumination generated by sectored (polygon) condenser plate[J]. Journal of Physics: Conference Series, 2009, 186: 012020.

[19] Vogt U, Reinspach J, Uhlén F, et al. Diffractive optics for laboratory sources to free electron lasers[J]. Journal of Physics: Conference Series, 2013, 463: 012001.

[20] Watanabe N, Hashizume J, Goto M, et al. Differential phase microscope and micro-tomography with a Foucault knife-edge scanning filter[J]. Journal of Physics: Conference Series, 2013, 463: 012011.

[21] Wong J. D'Sa D, Foley M, et al. NanoXCT: a novel technique to probe the internal architecture of pharmaceutical particles[J]. Pharmaceutical Research, 2014, 31(11): 3085-3094.

[22] 杨云昊. 硬X射线显微和纳米CT技术在细胞成像中的应用[D]. 合肥: 中国科学技术大学, 2010.

    Yang YH. Study on the applications of hard X-ray microscopy and nano-CT in cellular imaging[D]. Hefei: University of Science and Technology of China, 2010.

[23] Stock S R, Bleuet P, Laloum D, et al. SEM-based system for 100nm X-ray tomography for the analysis of porous silicon[J]. Proceedings of SPIE, 2014, 9212: 92120Z.

[24] Mohacsi I, Vartiainen I, Rösner B, et al. Interlaced zone plate optics for hard X-ray imaging in the 10 nm range[J]. Scientific Reports, 2017, 7: 43624.

[25] Krüger S P, Neubauer H, Bartels M, et al. Sub-10 nm beam confinement by X-ray waveguides: design, fabrication and characterization of optical properties[J]. Journal of Synchrotron Radiation, 2012, 19(2): 227-236.

[26] 陶芬, 王玉丹, 任玉琦, 等. X射线纳米成像单毛细管椭球镜的设计与检测[J]. 光学学报, 2017, 37(10): 1034002.

    Tao F, Wang Y D, Ren Y Q, et al. Design and detection of ellipsoidal mono-capillary for X-ray nano-imaging[J]. Acta Optica Sinica, 2017, 37(10): 1034002.

[27] Uesugi K, Hoshino M, Yagi N. Comparison of lens- and fiber-coupled CCD detectors for X-ray computed tomography[J]. Journal of Synchrotron Radiation, 2011, 18(2): 217-223.

[28] Chkhalo N I, Pestov A E, Salashchenko N N, et al. Sub-micrometer resolution proximity X-ray microscope with digital image registration[J]. Review of Scientific Instruments, 2015, 86(6): 063701.

[29] Stampanoni M, Wyss P, Abela R, et al. X-ray tomographic microscopy at the Swiss Light Source[J]. Proceedings of SPIE, 2002, 4503: 42-53.

[30] Koch A, Raven C, Spanne P, et al. X-ray imaging with submicrometer resolution employing transparent luminescent screens[J]. Journal of the Optical Society of America A, 1998, 15(7): 1940-1951.

[31] Engblom C, Langlois F, et al. Interferometric characterization of rotation stages for X-ray nanotomography[J]. Review of Scientific Instruments, 2017, 88(5): 053703.

[32] Nicola M, Hrouzek M, Renier M, et al. Towards a 10 nm run-out rotation axis [EB/OL]. [ 2019- 12- 31]. http://medsi2006spring8orjp/proc/43pdf, 2006.

[33] Kim J, Lauer K, Yan H, et al. Compact prototype apparatus for reducing the circle of confusion down to 40 nm for X-ray nanotomography[J]. Review of Scientific Instruments, 2013, 84(3): 035006.

[34] Xu W H, Lauer K, Chu Y, et al. A high-precision instrument for mapping of rotational errors in rotary stages[J]. Journal of Synchrotron Radiation, 2014, 21(6): 1367-1369.

[35] Zhao J T, Hu X D, Zou J, et al. Method for correction of rotation errors in Micro-CT System[J]. Nuclear Instruments and Methods in Physics Research Section A, 2016, 816: 149-159.

[36] Fu J, Li C, Liu Z Z. Analysis and calibration of stage axial vibration for synchrotron radiation nanoscale computed tomography[J]. Analytical and Bioanalytical Chemistry, 2015, 407(25): 7647-7655.

[37] Fu J, Li C, Liu Z Z. Analysis and correction of dynamic geometric misalignment for nano-scale computed tomography at BSRF[J]. PLoS One, 2015, 10(10): e0141682.

[38] Attwood D. Nanotomography comes of age[J]. Nature, 2006, 442(7103): 642-643.

[39] Chao W L, Harteneck B D, Liddle J A, et al. Soft X-ray microscopy at a spatial resolution better than 15 nm[J]. Nature, 2005, 435(7046): 1210-1213.

[40] Li KN, Wojcik MJ, DivanR, et al., Nanotechnology and Microelectronics:Materials, Processing, Measurement, Phenomena, 2017, 35(6): 06G901.

[41] MoldovanN, DivanR, Zeng HJ, et al., Surfaces, Films, 2018, 36(1): 01A124.

[42] Takeuchi A, Uesugi K, Suzuki Y. Improvement of quantitative performance of imaging X-ray microscope by reduction of edge-enhancement effect[J]. Journal of Physics: Conference Series, 2017, 849: 012055.

[43] ParfeniukasK. High-aspect ratio nanofabrication for hard X-ray zone plates[D]. Stockholm: KTH Royal Institute of Technology, 2018.

[44] Takeuchi A, Uesugi K, Suzuki Y, et al. Fresnel zone plate with apodized aperture for hard X-ray Gaussian beam optics[J]. Journal of Synchrotron Radiation, 2017, 24(3): 586-594.

[45] Chen T Y, Chen Y T, Wang C L, et al. Full-field microimaging with 8 keV X-rays achieves a spatial resolutions better than 20 nm[J]. Optics Express, 2011, 19(21): 19919-19924.

[46] Markus O, Matthias B, Döring F, et al. Two-dimensional sub-5-nm hard X-ray focusing with MZP[J]. Proceedings of SPIE, 2013, 8848: 884802.

[47] Osterhoff M, Eberl C, Döring F, et al. Towards multi-order hard X-ray imaging with multilayer zone plates[J]. Journal of Applied Crystallography, 2015, 48(1): 116-124.

[48] Liu C A, Conley R. MacRander A T, et al. Depth-graded multilayers for application in transmission geometry as linear zone plates[J]. Journal of Applied Physics, 2005, 98(11): 113519.

[49] Kang H C, Maser J, Stephenson G B, et al. Nanometer linear focusing of hard X rays by a multilayer Laue lens[J]. Physical Review Letters, 2006, 96(12): 127401.

[50] KoyamaT, TsujiT, TakanoH, et al. Development of multilayer laue lenses; (2) circular type[C]. AIP Conference Proceedings, 2011, 1365( 1): 100- 103.

[51] MimuraH, KimuraT, YokoyamaH, et al. Development of an adaptive optical system for sub-10-nm focusing of synchrotron radiation hard X-rays[C]. AIP Conference Proceedings, 2011, 1365( 1): 13- 20.

[52] Huang X J, Yan H F, Nazaretski E, et al. 11 nm hard X-ray focus from a large-aperture multilayer Laue lens[J]. Scientific Reports, 2013, 3: 3562.

[53] Huang X J, Conley R, Bouet N, et al. Achieving hard X-ray nanofocusing using a wedged multilayer Laue lens[J]. Optics Express, 2015, 23(10): 12496-12507.

[54] Nazaretski E, Xu W, Bouet N, et al. Development and characterization of monolithic multilayer Laue lens nanofocusing optics[J]. Applied Physics Letters, 2016, 108(26): 261102.

[55] Li Y L, Beck R, Huang T, et al. Scatterless hybrid metal-single-crystal slit for small-angle X-ray scattering and high-resolution X-ray diffraction[J]. Journal of Applied Crystallography, 2008, 41(6): 1134-1139.

[56] Cai Z H, Rodrigues W, Ilinski P, et al. Synchrotron X-ray microdiffraction diagnostics of multilayer optoelectronic devices[J]. Applied Physics Letters, 1999, 75(1): 100-102.

[57] Chon K S, Juhng S K, Yoon K H. Design study of hard X-ray tomography system to obtain a spatial resolution of 100 nm[J]. Current Applied Physics, 2012, 12(1): 134-140.

[58] Zeng X H, Duewer F, Feser M, et al. Ellipsoidal and parabolic glass capillaries as condensers for X-ray microscopes[J]. Applied Optics, 2008, 47(13): 2376-2381.

[59] Wang Y D, Ren Y Q, Zhou G Z, et al. Equally sloped tomography based X-ray full-field nano-CT at Shanghai Synchrotron Radiation Facility[J]. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2018, 896: 108-112.

[60] Jarre A, Fuhse C, Ollinger C, et al. Two-dimensional hard X-ray beam compression by combined focusing and waveguide optics[J]. Physical Review Letters, 2005, 94(7): 074801.

[61] Kirkpatrick P, Baez A V. Formation of optical images by X-rays[J]. Journal of the Optical Society of America, 1948, 38(9): 766.

[62] Mimura H, Handa S, Kimura T, et al. Breaking the 10 nm barrier in hard-X-ray focusing[J]. Nature Physics, 2010, 6(2): 122-125.

[63] Snigirev A, Kohn V, Snigireva I, et al. Focusing high-energy X-rays by compound refractive lenses[J]. Applied Optics, 1998, 37(4): 653-662.

[64] Ladislav P N, Yury D, Vaclav J, et al. X-ray imaging with compound refractive lens and microfocus X-ray tube[J]. Proceedings of SPIE, 2008, 7077: 70770H.

[65] Schroer C G, Kurapova O, Patommel J, et al. Hard X-ray nanoprobe based on refractive X-ray lenses[J]. Applied Physics Letters, 2005, 87(12): 124103.

[66] Brancewicz M, Itou M, Sakurai Y, et al. High transmission Ni compound refractive lens for high energy X-rays[J]. Review of Scientific Instruments, 2016, 87(8): 085106.

[67] Lengeler B, Tümmler J, Snigirev A, et al. Transmission and gain of singly and doubly focusing refractive X-ray lenses[J]. Journal of Applied Physics, 1998, 84(11): 5855-5861.

[68] Piestrup M A, Cremer J T, Beguiristain H R, et al. Two-dimensional X-ray focusing from compound lenses made of plastic[J]. Review of Scientific Instruments, 2000, 71(12): 4375.

[69] Aristov V, Grigoriev M, Kuznetsov S, et al. X-ray refractive planar lens with minimized absorption[J]. Applied Physics Letters, 2000, 77(24): 4058-4060.

[70] LengelerB, Schroer CG, BennerB, et al. and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and AssociatedEquipment, 2001, 467/468: 944- 950.

[71] Aaron S, Kenneth E, Ashley T. Kinoform lenses: toward nanometer resolution[J]. Proceedings of SPIE, 2005, 6002: 600210.

[72] Karvinen P, Grolimund D, Willimann M, et al. Kinoform diffractive lenses for efficient nano-focusing of hard X-rays[J]. Optics Express, 2014, 22(14): 16676-16685.

[73] Chen Y T, Lo T N, Chu Y S, et al. Full-field hard X-ray microscopy below 30 nm: a challenging nanofabrication achievement[J]. Nanotechnology, 2008, 19(39): 395302.

[74] Mohacsi I, Vartiainen I, Guizar-Sicairos M, et al. High resolution double-sided diffractive optics for hard X-ray microscopy[J]. Optics Express, 2015, 23(2): 776-786.

[75] Takemoto K, Usui K, Ohigashi T, et al. Improvement of cryogenic 3-dimensional observation system of soft X-ray microscope at the SR center of Ritsumeikan University[J]. Journal of Physics: Conference Series, 2013, 463: 012009.

[76] WeitkampT. High-resolution X-ray imaging and tomography at the ESRF beamline ID 22[C]. AIP Conference Proceedings, 2000, 507( 1): 424- 429.

[77] Chu Y S. Yi J M, de Carlo F, et al. Hard-X-ray microscopy with Fresnel zone plates reaches 40 nm Rayleigh resolution[J]. Applied Physics Letters, 2008, 92(10): 103119.

[78] Takeuchi A, Uesugi K, Takano H, et al. Submicrometer-resolution three-dimensional imaging with hard X-ray imaging microtomography[J]. Review of Scientific Instruments, 2002, 73(12): 4246-4249.

[79] Yoshio S, Takeuchi A, Terada Y, et al. Development of large-field high-resolution hard X-ray imaging microscopy and microtomography with Fresnel zone plate objective[J]. Proceedings of SPIE, 2013, 8851: 885109.

[80] Pogany A. A small step to higher resolution[J]. Nature Physics, 2006, 2(10): 657-658.

[81] Ice G E, Budai J D. Pang J W L. The race to X-ray microbeam and nanobeam science[J]. Science, 2011, 334(6060): 1234-1239.

[82] Kohmura Y, Awaji M, Suzuki Y, et al. X-ray focusing test and X-ray imaging test by a microcapillary X-ray lens at an undulator beamline[J]. Review of Scientific Instruments, 1999, 70(11): 4161-4167.

[83] Schroer C G, Kuhlmann M, Hunger U T, et al. Nanofocusing parabolic refractive X-ray lenses[J]. Applied Physics Letters, 2003, 82(9): 1485-1487.

[84] SnigirevaI, Vaughan G B M, Snigirev A, et al. High-energy nanoscale-resolution X-ray microscopy based on refractive optics on a long beamline[C]. AIP Conference Proceedings, 2011, 1365( 1): 188- 191.

[85] Lengeler B, Schroer C G, Richwin M, et al. A microscope for hard X-rays based on parabolic compound refractive lenses[J]. Applied Physics Letters, 1999, 74(26): 3924-3926.

[86] Liao K L, Hong Y L, Sheng W F. Optimized short kinoform lenses for hard X-ray nano-focusing[J]. Optics Communications, 2015, 339: 53-60.

[87] 钟长游. 高能X射线复合折射透镜设计和应用[D]. 北京: 中国科学院大学, 2018.

    Zhong CY. Design and application of high energy X-ray compound refractive lens[D]. Beijing: University of Chinese Academy of Sciences, 2018.

[88] Seiboth F, Wittwer F, Scholz M, et al. Nanofocusing with aberration-corrected rotationally parabolic refractive X-ray lenses[J]. Journal of Synchrotron Radiation, 2018, 25(1): 108-115.

[89] Kohn V G. Effective aperture of X-ray compound refractive lenses[J]. Journal of Synchrotron Radiation, 2017, 24(3): 609-614.

[90] Gasilov S, dos Santos Rolo T, Mittone A, et al. Generalized pupil function of a compound X-ray refractive lens[J]. Optics Express, 2017, 25(21): 25090-25097.

[91] Huang CC, Mu BZ, Wang ZS, et al. Imaging properties of a spherical compound refractive X-ray lens[C]. Proceedings of SPIE, 2009, 7360: 736006.

[92] Korytár D, Cecilia A, et al. High-resolution high-efficiency X-ray imaging system based on the in-line Bragg magnifier and the Medipix detector[J]. Journal of Synchrotron Radiation, 2013, 20(1): 153-159.

[93] Hirano K, Yamashita Y, Takahashi Y, et al. Development of variable-magnification X-ray Bragg optics[J]. Journal of Synchrotron Radiation, 2015, 22(4): 956-960.

[94] Švéda L, Cecilia A, et al. X-ray Bragg magnifier microscope as a linear shift invariant imaging system: image formation and phase retrieval[J]. Optics Express, 2014, 22(18): 21508-21520.

[95] Hrivňak S. Uli n J, Mike L, et al. Single-distance phase retrieval algorithm for Bragg Magnifier microscope[J]. Optics Express, 2016, 24(24): 27753-27762.

[96] Stampanoni M, Borchert G, Abela R, et al. Bragg magnifier: a detector for submicrometer X-ray computer tomography[J]. Journal of Applied Physics, 2002, 92(12): 7630-7635.

[97] Dabagov S B. Wave theory of X-ray scattering in capillary structures[J]. X-Ray Spectrometry, 2003, 32(3): 179-185.

[98] 丰丙刚. 毛细管聚焦特性及其X射线荧光与纳米全场成像应用研究[D]. 上海: 中国科学院研究生院( 上海应用物理研究所), 2017.

    Feng BG. The focusing characteristics of capillary and its applications on the fluorescence imaging and full-field X-ray nano-imaging[D]. Shanghai: Shanghai Institute of Applied Physics,Chinese Academy of Sciences, 2017.

[99] Emoto T, Sato Y, Konishi Y, et al. Development and applications of grazing exit micro X-ray fluorescence instrument using a polycapillary X-ray lens[J]. Spectrochimica Acta Part B: Atomic Spectroscopy, 2004, 59(8): 1291-1294.

[100] Yang J, Li Y D, Wang X Y, et al. Simulation and application of micro X-ray fluorescence based on an ellipsoidal capillary[J]. Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions With Materials and Atoms, 2017, 401: 25-28.

[101] Mazuritskiy M I, Lerer A M. Spatial distribution of channeling long-wavelength X rays at the output of polycapillary structures[J]. JETP Letters, 2015, 102(7): 483-486.

[102] Yamanashi M, Kometani N, Tsuji K. Preliminary experiment of X-ray diffraction imaging[J]. Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions With Materials and Atoms, 2015, 355: 272-275.

[103] Mazuritskiy M I, Lerer A M, Makhno P V. Anomalous scattering, transport, and spatial distribution of X-ray fluorescence at the exit of polycapillary structures[J]. Journal of Experimental and Theoretical Physics, 2016, 123(6): 942-949.

[104] Chen J P, Wang J Y, Zou J, et al. Polycapillary coupled X-ray digital radiation imaging system: Feasibility analysis[J]. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2017, 870: 19-24.

[105] Chen Y T, Chen T Y, Yi J, et al. Hard X-ray Zernike microscopy reaches 30 nm resolution[J]. Optics Letters, 2011, 36(7): 1269-1271.

[106] Holzner C, Feser M, Vogt S, et al. Zernike phase contrast in scanning microscopy with X-rays[J]. Nature Physics, 2010, 6(11): 883-887.

[107] Bai B, Zhu R K, Wu S T, et al. Multi-scale method of Nano(Micro)-CT study on microscopic pore structure of tight sandstone of Yanchang Formation, Ordos Basin[J]. Petroleum Exploration and Development, 2013, 40(3): 354-358.

[108] Bailey J J. Heenan T M M, Finegan D P, et al. Laser-preparation of geometrically optimised samples for X-ray nano-CT[J]. Journal of Microscopy, 2017, 267(3): 384-396.

[109] Do M, Isaacson S A. McDermott G, et al. Imaging and characterizing cells using tomography[J]. Archives of Biochemistry and Biophysics, 2015, 581: 111-121.

[110] Fischer P. X-ray imaging of magnetic structures[J]. IEEE Transactions on Magnetics, 2015, 51(2): 1-31.

[111] Gureyev T E. Nesterets Y I, de Hoog F, et al. Duality between noise and spatial resolution in linear systems[J]. Optics Express, 2014, 22(8): 9087-9094.

[112] Hasegawa T, Hanada T, Yorozu A, et al. Microfocus X-ray imaging of the internal geometry of brachytherapy seeds[J]. Applied Radiation and Isotopes, 2014, 86: 13-20.

[113] Huang X Z, Li N, Wang D J, et al. Quantitative three-dimensional analysis of poly (lactic-co-glycolic acid) microsphere using hard X-ray nano-tomography revealed correlation between structural parameters and drug burst release[J]. Journal of Pharmaceutical and Biomedical Analysis, 2015, 112: 43-49.

[114] Kim C, Zuo Z L, Finger H, et al. Soft X-ray-assisted detection method for airborne molecular contaminations (AMCs)[J]. Journal of Nanoparticle Research, 2015, 17(3): 126.

[115] Lim C, Yan B, Yin L L, et al. Simulation of diffusion-induced stress using reconstructed electrodes particle structures generated by micro/nano-CT[J]. Electrochimica Acta, 2012, 75: 279-287.

[116] Litster S, Epting W K, Wargo E A, et al. Morphological analyses of polymer electrolyte fuel cell electrodes with nano-scale computed tomography imaging[J]. Fuel Cells, 2013, 13: 935-945.

[117] Vogt S, Schneider G, Steuernagel A, et al. X-ray microscopic studies of the drosophila dosage compensation complex[J]. Journal of Structural Biology, 2000, 132(2): 123-132.

[118] Momose A, Kawamoto S, Koyama I, et al. Demonstration of X-ray Talbot interferometry[J]. Japanese Journal of Applied Physics, 2003, 42(7B): L866-L868.

[119] Hauke C, Anton G, Hellbach K, et al. Enhanced reconstruction algorithm for moiré artifact suppression in Talbot-Lau X-ray imaging[J]. Physics in Medicine & Biology, 2018, 63(13): 135018.

[120] 王雅丽, 史祎诗, 李拓, 等. 可见光域叠层成像中照明光束的关键参量研究[J]. 物理学报, 2013, 62(6): 201-210.

    Wang Y L, Wang Y L, Li T, et al. Research on the key parameters of illuminating beam for imaging via ptychography in visible light band?[J]. Acta Physica Sinica, 2013, 62(6): 201-210.

[121] Giewekemeyer K, Thibault P, Kalbfleisch S, et al. Quantitative biological imaging by ptychographic X-ray diffraction microscopy[J]. Proceedings of the National Academy of Sciences of the United States of America, 2010, 107(2): 529-534.

[122] Holler M, Diaz A, Guizar-Sicairos M, et al. X-ray ptychographic computed tomography at 16 nm isotropic 3D resolution[J]. Scientific Reports, 2015, 4: 3857.

[123] Dierolf M, Menzel A, Thibault P, et al. Ptychographic X-ray computed tomography at the nanoscale[J]. Nature, 2010, 467(7314): 436-439.

[124] Kruth J P, Bartscher M, Carmignato S, et al. Computed tomography for dimensional metrology[J]. CIRP Annals, 2011, 60(2): 821-842.

[125] LazaroD, LegoupilS, BlokkeelG, et al. Metrology of steel micro-nozzles using X-ray microtomography[C]//Proceedings of the DIR 2007 - International Symposium on Digital industrial Radiology and Computed Tomography, 2007: 25- 27.

[126] Villarraga-Gómez H, Herazo E L, et al. X-ray computed tomography: from medical imaging to dimensional metrology[J]. Precision Engineering, 2019, 60: 544-569.

[127] Schmitt R, Niggemann C. Uncertainty in measurement for X-ray-computed tomography using calibrated work pieces[J]. Measurement Science and Technology, 2010, 21(5): 054008.

[128] Villarraga-Gómez H, Lee C, Smith S T. Dimensional metrology with X-ray CT: a comparison with CMM measurements on internal features and compliant structures[J]. Precision Engineering, 2018, 51: 291-307.

[129] 唐天旭, 段晓礁, 周志政, 等. 基于散射校正板的锥束微纳CT系统的散射校正[J]. 光学学报, 2019, 39(8): 0834001.

    Tang T X, Duan X J, Zhou Z Z, et al. Scatter correction based on beam stop array for cone-beam micro-computed tomography[J]. Acta Optica Sinica, 2019, 39(8): 0834001.

[130] Dudak J, Karch J, Holcova K, et al. X-ray imaging with sub-micron resolution using large-area photon counting detectors Timepix[J]. Journal of Instrumentation, 2017, 12(12): C12024.

[131] 李文杰. 纳米CT三维图像处理分析方法及其应用的研究[D]. 合肥: 中国科学技术大学, 2011: 20- 28.

    Li WJ. Study on the application of 3D image processing and analysis for nano-CT[D]. Hefei: University of Science and Technology of China, 2011: 20- 28.

[132] 朱佩平, 何其利, 廖可梁, 等. 具有三种定量成像机制的纳米CT三维成像方法: CN201711292776. 5[P].2017-12-08.

    Zhu PP, He QL, Liao KL, et al. 2017-12-08.

吕寒玉, 邹晶, 赵金涛, 胡晓东. 纳米计算机断层扫描成像技术进展综述[J]. 激光与光电子学进展, 2020, 57(14): 140001. Hanyu Lü, Jing Zou, Jintao Zhao, Xiaodong Hu. Review on Development of Nano-Computed Tomography Imaging Technology[J]. Laser & Optoelectronics Progress, 2020, 57(14): 140001.

本文已被 6 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!