中国激光, 2024, 51 (3): 0307102, 网络出版: 2024-01-19  

采用光子计数测量的高灵敏度锥束XLCT

High‑Sensitivity Cone‑Beam XLCT Using Photon Counting Measurements
作者单位
1 天津大学精密仪器与光电子工程学院,天津 300072
2 天津市生物医学检测技术与仪器重点实验室,天津 300072
摘要
X射线激发发光断层成像(XLCT)是一种新兴的混合成像技术,该技术可同时获得目标体结构信息和功能信息。然而,窄束XLCT重建图像虽然具有高空间分辨率,但X射线利用率较低,数据采集时间过长;锥束XLCT系统提高了X射线利用率,缩短了数据采集时间,但有限的探测角度导致图像重建质量相对较低。为解决上述问题,提高X射线利用率,高效实现多角度高灵敏数据采集,提出了一种基于光子计数测量的锥束XLCT系统,并通过仿体实验对其性能进行了验证。仿体实验结果表明,6个角度锥束照射模式下重建图像的相似度系数(DICE)可达50%,系统保真度(SF)和重建浓度误差(RCE)达到0.7以上,可实现3 mg·mL-1以上的荧光物质质量浓度差距的重建。
Abstract
Objective

X-ray luminescence computed tomography (XLCT) technology uses X-ray excitation to stimulate specific luminescent materials at the nanoscale, termed phosphor nanoparticles (PNPs), to produce near-infrared light. Photodetectors then capture the emitted near-infrared light signals from these excited PNPs. Through suitable algorithms, the distribution of PNPs within biological tissues can be visualized. This method allows for structural and functional insights into biological tissues, showing great potential for advancement. There are two main types of XLCT systems: narrow-beam and cone-beam. The narrow-beam XLCT system exhibits higher spatial resolution, albeit at the cost of lower X-ray utilization efficiency. This inefficiency results in extended imaging times, limiting its potential for clinical use. Conversely, the cone-beam XLCT system improves X-ray efficiency and shortens detection time. However, the quality of the reconstructed images tends to be lower due to detection angle limitations. To overcome these challenges, there is a need for an innovative XLCT system that realizes rapid and highly sensitive data collection while also maximizing the use of X-ray technology. By addressing these issues, the clinical limitations of XLCT can be reduced to pave the way for its further development, thereby unlocking a plethora of possibilities.

Methods

This study introduces a new cone-beam XLCT system based on photon-counting measurements, complemented by an associated reconstruction method. Through the synergistic collaboration between the field-programmable gate array (FPGA) based sub-sampling unit and upper-level control unit, the system realizes automated multi-channel measurements. This integration shortens data acquisition time, boosts experimental efficiency, and mitigates the risks associated with X-ray exposure. After the completion of system implementation, we conduct experimental validation of the system and methodology. Specifically, a fabricated phantom is subjected to multi-angle projection measurements using the established system, and image reconstruction and evaluation are performed using the Tikhonov reconstruction algorithm.

Results and Discussions

The results of the dual target phantom experiment indicate that under the conditions of a cylindrical phantom radius of 40 mm, target radius of 6 mm, and distance of 14 mm from the dual target phantom (Fig.2), the similarity coefficient (DICE) of the reconstructed image of the dual target phantom exceeds 50% under six-angle cone-beam X-ray irradiation. Furthermore, the system fidelity (SF) exceeds 0.7 (Table 1). In the phantom experiment of dual targets with different concentrations, the system proposed in this study effectively distinguishes dual targets with a mass concentration difference of more than 3 mg/mL. The DICE of the reconstruction image maintains over 50%, SF remains over 0.7, and reconstruction concentration error (RCE) is also over 0.7 (Table 2). These phantom experiment results confirm the good fidelity and resolution capability of the proposed system. Nevertheless, numerous factors potentially degrade the experimental outcomes, such as the attenuation and scattering of X-ray beams in the XLCT system, the physical and chemical composition of the target body, or even uneven concentration distribution. Additionally, artifacts appear in the reconstructed images. In the future, our research will focus on optimizing algorithms and reducing noise to enhance the application of cone-beam XLCT for in vivo experiments.

Conclusions

This study comprehensively considers the advantages and disadvantages of two imaging methods in XLCT and proposes a photon-counting-based multi-channel cone-beam XLCT system. The system automation for multi-angle measurements is realized via FPGA and host computer interaction. Specifically, multi-angle cone-beam irradiation reduces data acquisition time, while photon-counting measurement enhances the system sensitivity. Furthermore, a phantom experiment is conducted to validate the effectiveness and practicality of the proposed system and algorithm. The results demonstrate a significant reduction in data acquisition time and an improvement in the utilization of X-rays.

韩景灏, 贾梦宇, 周仲兴, 高峰. 采用光子计数测量的高灵敏度锥束XLCT[J]. 中国激光, 2024, 51(3): 0307102. Jinghao Han, Mengyu Jia, Zhongxing Zhou, Feng Gao. High‑Sensitivity Cone‑Beam XLCT Using Photon Counting Measurements[J]. Chinese Journal of Lasers, 2024, 51(3): 0307102.

引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!